Cycles |
Standard: sterilizes most general surgical instruments
Advanced: sterilizes single-channel rigid endoscopes or up to 1 single-channel flexible endoscope |
Time |
Standard: 28 minutes Advanced: 38 minutes |
Lumen claims |
Standard: single-channel stainless steel lumens diameter: ≥1 mm; length: ≤150 mm OR diameter: ≥2 mm; length: ≤400 mm
Advanced: single-channel stainless steel lumens diameter: ≥1 mm; length: ≤500 mm OR single-channel PE/PTFE flexible endoscopes diameter: ≥1 mm; length: ≤850 mm |
Cycle temperature |
47°C–56°C |
Sterilant |
Hydrogen peroxide |
Sterilant delivery |
Delivered in closed system in cassettes with automatic detection of expiration date:
Standard/Advanced: 3.6 mL per cycle (double kill) (2 ampules, each ampule is 1.8 mL) |
Used cassette disposal |
Automatic and touchless ejection into cassette disposal container |
Peroxide residual breakdown |
Gas plasma technology breaks down H2O2 to safe elements of water and oxygen |
Configurations |
Single door |
System dimensions (maximum) |
System Height: 930 mm Width: 554 mm Depth: 792 mm
System + Cart Height: 1673 mm Width: 576 mm Depth: 792 mm |
Chamber total volume |
51.3 L |
Chamber dimensions |
System Height: 230.1 mm Width: 370.1 mm Depth: 660.9 mm |
Chamber shape |
Rectangular |
Shelf info |
Two-tiered shelf: Width: 312.8 mm Depth: 600.3 mm |
Mobility (Cart) |
4 casters (all locking) |
System weight |
System: 127.2 kg Cart: 65.3 kg |
User interface |
Touchscreen technology: projected capacitive touch Resolution: 800 x 600 pixels |
Accessory devices |
• Barcode reader • External drives: USB that allows data upload and download |
Standards/Compliance |
ISO 14937 |
Sterilization process |
Terminal-sterilization, double-kill cycle to provide a Sterility Assurance Level (SAL) of 10-6; 2 injections and identical plasma phases |
Delivered sterilant concentration |
58%–59.5% |
Sterilization cycle monitoring |
Critical system parameters monitored with on-board sensors, biological indicators, chemical indicators, IMS (independent monitoring system) available |
H2O2 concentration continuous monitoring |
Monitoring using UV sensor within the chamber |
Heat generation |
Maximum: 1012 BTU/hour |
Electrical power specifications |
One dedicated single phase 50/60 Hz, voltage +/- 10%,
2+ ground outlet per the following countries:
• US/Canada: 120 VAC, 16A (recommended receptacle: NEMA 5-20R)
• Japan: 200 VAC, 10A
• International: 200-240 VAC, 10A; 110-120 VAC, 16A |
Service space and installation requirements (Space can be reduced if the system can be moved for servicing.) |
Top: 610 mm minimum
Left side:
• Table installed: 60 mm minimum from sterilizer side
• Cart installed: 100 mm minimum from cart side
Right side:
• Table installed: 610 mm minimum from sterilizer side
(Objects placed within this region must be moved during service. Area can be divided up into
space on left and right sides.)
• Cart installed: 100 mm minimum from cart side
Back side: 25.4 mm minimum
Front: 1000 mm minimum from front of sterilizer |
Countertop/table specifications |
Structure must be able to support a minimum of 182 kg, and have a minimum depth of 584 mm |
Temperature |
18°C–35°C |
Humidity |
10%–85% RH (noncondensing) |
System performance data and reports |
Cycle history, full 1-second data files, and reports available via ASP ACCESS™ Technology |
Network connectivity |
Communication protocol for Instrument Tracking Systems (ITS) available via ASP ACCESS™ Technology |
Data recording |
• Electronic data storage up to 50 cycles • Internal printer for manual recordkeeping
• Full electronic cycle data and reports via ASP ACCESS™ Technology |
Code |
Product Description |
10033-004 |
STERRAD NX® Sterilization System with ALLClear™ Technology, STANDARD & ADVANCED cycles, includes installation |
10311 |
STERRAD NX® Sterilization System with ALLClear™ Technology Cart |
10133 |
STERRAD NX® System Cassette (5 cassettes/case) |
113617-01 |
STERRAD® Systems Bar Code Scanner Kit |
10306 |
Cassette disposal box |
10305 |
Thermal printer paper |